rescale slope and rescale intercept

5YrsLaterDBA picture 5YrsLaterDBA · Apr 17, 2012 · Viewed 18.2k times · Source

I have few questions about the rescale slope and rescale intercept in CT DICOM:

  1. Are they used to transfer original data from the scanner to Hounsfield data set, in which water is 0 and air is -1000?
  2. I am in the image display group. How can I know the rescale slope and the rescale intercept values?
  3. What's the exact meaning of the rescale slope and the rescale intercept? How are they determined?

Answer

Paolo Brandoli picture Paolo Brandoli · Apr 17, 2012

The rescale slope and rescale intercept allow to transform the pixel values to HU or other units, as specified in the tag 0028,1054.

For CT images, the unit should be HU (Hounsfield) and the default value is indeed HU when the tag 0028,1054 is not present. However, the tag may be present and may specify a different unit (OD=optical density, US=unspecified).

The rescale slope and intercept are determined by the manufacturer of the hardware.

If the transformation from original pixel values to Hounsfield or Optical density is not linear, then a LUT is applied.

Check the part 3 of the standard C.11 for more detailed information, and also this answer Window width and center calculation of DICOM image